| Catalog Number |
22-4035 |
| Medical Device Problem Codes |
Break (1069); Difficult to Remove (1528); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a shoulder arthroscopy, when assembling the clamp of the firstpass suture passer, the rack part of it was damaged (nonetheless, the user was able to assemble and use it).Additionally, the wire was not bringing, it was pulled out with a grasper.The procedure was successfully completed with a delay of less than 30 minutes using the same device.There was no damage to the patient's health.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.However, an image evaluation was performed and found the complaint device, the upper jaw is missing a piece.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include excessive force, tissue thickness, damage or debris on the device tip between passes.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.Corrected data: h6: medical device problem code.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the reported device was received for evaluation.A visual evaluation showed one device was returned but not in any original packaging.The end of the upper jaw is missing.Tape or plastic is wrapped around the suture capture opening.There are indentions from another instrument being used on the upper jaw.Bio debris is present.A functional evaluation showed the suture capture loaded properly but would not unload using the #4 unloader due to the broken jaw.The needle loaded, deployed, and unloaded as intended.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.An analysis of the customer provided image finds the complaint device, the upper jaw is missing a piece.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risk and leading to unintended consequences, and excessive force, tissue thickness, damage or debris on the device tip between passes.Based on this investigation, there is no evidence to suggest a design, material or manufacturing issue.Therefore, the need for corrective action is not indicated.
|
| |
|
Additional Manufacturer Narrative
|
|
H2: corrected data on: d4 (lot number, expiration date (should be left in blank)) and h4.
|
| |
|