MAUDE Adverse Event Report: ARTHROCARE CORPORATION FIRSTPASS SUTURE PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORPORATION FIRSTPASS SUTURE PASSER Back to Search Results
Catalog Number 22-4035
Medical Device Problem Codes Break (1069); Difficult to Remove (1528); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during a shoulder arthroscopy, when assembling the clamp of the firstpass suture passer, the rack part of it was damaged (nonetheless, the user was able to assemble and use it).Additionally, the wire was not bringing, it was pulled out with a grasper.The procedure was successfully completed with a delay of less than 30 minutes using the same device.There was no damage to the patient's health.No further complications were reported.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.However, an image evaluation was performed and found the complaint device, the upper jaw is missing a piece.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include excessive force, tissue thickness, damage or debris on the device tip between passes.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.Corrected data: h6: medical device problem code.
 
Additional Manufacturer Narrative
H3, h6: the reported device was received for evaluation.A visual evaluation showed one device was returned but not in any original packaging.The end of the upper jaw is missing.Tape or plastic is wrapped around the suture capture opening.There are indentions from another instrument being used on the upper jaw.Bio debris is present.A functional evaluation showed the suture capture loaded properly but would not unload using the #4 unloader due to the broken jaw.The needle loaded, deployed, and unloaded as intended.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.An analysis of the customer provided image finds the complaint device, the upper jaw is missing a piece.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risk and leading to unintended consequences, and excessive force, tissue thickness, damage or debris on the device tip between passes.Based on this investigation, there is no evidence to suggest a design, material or manufacturing issue.Therefore, the need for corrective action is not indicated.
 
Additional Manufacturer Narrative
H2: corrected data on: d4 (lot number, expiration date (should be left in blank)) and h4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIRSTPASS SUTURE PASSER
Common Device Name
PASSER
Manufacturer (Section D)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18682345
Report Number3006524618-2024-00061
Device Sequence Number17690844
Product Code HWQ
UDI-Device Identifier00817470000160
UDI-Public00817470000160
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number22-4035
Device Lot NumberK5001BB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2024
Supplement Date Received by Manufacturer04/04/2024
09/09/2024
09/09/2024
Initial Report FDA Received Date02/09/2024
Supplement Report FDA Received Date04/05/2024
09/10/2024
09/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/20/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
-
-