| Model Number |
DST0856517 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the correct application, the hemostasis valve is leaking at the side of the housing.The patient is fine.There was no danger to the patient at any time.There was only a delay in the operation of approx.10 minutes due to the exchange for a correct destino twist.The replacement product was the identical destino twist sheath.The operation was performed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion not yet available, evaluation in process.A follow-up report will be submitted as soon as the investigation is complete.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|
| |
|
Additional Manufacturer Narrative
|
|
One 8.5f destino twist steerable guiding sheath was returned under rga# rg24032/ca with the dilator.There were no other accessories.Traces of blood were found on and inside the sheath and dilator.According to the event description summary, the hemostasis valve was leaking at the side of the housing.Upon evaluation of the returned sheath, it was found that the hemostatic valve was torn when viewed under magnification.The sheath was manually leak tested with the syringe connected to the stopcock and the tip occluded.The sheath did not exhibit leakage when tested using this method.The sheath was then tested using the iso 11070 for liquid leakage test method.The specification for valve leakage is 38kpa.The device was occluded at the distal tip and pressurized to approximately 35kpa when the hemostatic valve started to leak from the tear.Per manufacturing procedure (non-peelable sheath final assembly) assembling the cap and seal visually inspect seals 100% before use.Carefully inspect seals to ensure that the helical center cut is present before assembly.Do not pull, bend or separate seals during inspection.If the seal is not properly cut, reject the seal and do not use for assembly.Notify the appropriate manufacturing supervisor before proceeding.Per procedure (destino steerable guiding sheath in-process and final inspection): the leak test is performed according to procedures based on available leak tester.The leak test is performed by manufacturing personnel 100% and is observed by quality assurance personnel at an aql level.Per ifu: wet the dilator shaft with sterile saline solution prior to insertion through the hemostatic valve.Note: any device/component inserted through the hemostatic valve of the sheath should be wet and placed through the center of the valve to prevent tearing of the seal and leakage.Returned device analysis revealed the sheath was within manufacturing specifications.There was a tear in the hemostatic valve that caused the sheath to leak and fail the iso leak test.The sheaths are leak tested 100% by manufacturing and observed by qa at an aql level in-process, prior to shipping to the customer.The potential cause of the tear/leak could be if the dilator (or other device) was not inserted through the center of the valve as per ifu.According to the device history record, the sheath passed all in-process and qa final inspection steps before shipping to the customer including visual, dimensional, mechanical and leak testing.No manufacturing defects were found.In conclusion, based on the information available at this time it cannot be confirmed that the product failed to meet requirements.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|
| |
|
Search Alerts/Recalls
|
|
|