| Model Number |
AU00T0 |
| Medical Device Problem Code |
Failure to Unfold or Unwrap (1669)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A facility representative reported that during an intraocular lens (iol) implantation procedure, the lens would not unfold.Lens was discarded no patient harm reported.No further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.Viscoelastic was not provided.It is unknown if a qualified product was used.The product investigation could not identify a root cause.It is unknown if a qualified viscoelastic was used.The ifu instructs: during device preparation and implantation of the iol with the preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.The account indicated the product was not available to evaluate.Follow up attempts were made.The account indicated there was no further information has been provided at this time.File will be reopened when new information is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|