Model Number MI1000 MED-EL CONCERT
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/06/2024
Type of Reportable Event Malfunction
Event or Problem Description
The user informed his parents upon returning from school, that he has lost his hearing through the implant.It is currently unknown whether this has been caused by trauma or accident.Re-implantation is considered.No date has been scheduled yet.
 
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The user informed his parents upon returning from school, that he has lost his hearing through the implant.It is currently unknown whether this has been caused by trauma or accident.Re-implantation is considered.No date has been scheduled yet.
 
Additional Manufacturer Narrative
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
 
Additional Manufacturer Narrative
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.The concerned device was explanted but has not been received for investigation yet.
 
Event or Problem Description
The user informed his parents upon returning from school, that he has lost his hearing through the implant.It is currently unknown whether this has been caused by trauma or accident.The user has been re-implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key18684951
Report Number9710014-2024-00132
Device Sequence Number15015179
Product Code MCM
UDI-Device Identifier09008737062682
UDI-Public(01)09008737062682
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/01/2021
Device Model NumberMI1000 MED-EL CONCERT
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2024
Supplement Date Received by Manufacturer02/07/2024
02/07/2024
Initial Report FDA Received Date02/12/2024
Supplement Report FDA Received Date04/16/2024
04/25/2024
Date Device Manufactured01/30/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age9 YR
Patient SexMale
-
-