| Model Number |
MI1000 MED-EL CONCERT |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user informed his parents upon returning from school, that he has lost his hearing through the implant.It is currently unknown whether this has been caused by trauma or accident.Re-implantation is considered.No date has been scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user informed his parents upon returning from school, that he has lost his hearing through the implant.It is currently unknown whether this has been caused by trauma or accident.Re-implantation is considered.No date has been scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.The concerned device was explanted but has not been received for investigation yet.
|
| |
|
Event or Problem Description
|
|
The user informed his parents upon returning from school, that he has lost his hearing through the implant.It is currently unknown whether this has been caused by trauma or accident.The user has been re-implanted.
|
| |
|
Search Alerts/Recalls
|