| Catalog Number |
183460 |
| Medical Device Problem Code |
Material Erosion (1214)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Joint Laxity (4526); Insufficient Information (4580); Joint effusion (4849)
|
| Date of Event |
01/17/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an initial total knee arthroplasty.Subsequently, six (6) years and eight (8) months post-implantation, the patient underwent revision surgery due to wear of the articular surface.The bearing was exchanged without reported complication.Due diligence is in progress for this event; to date no further information has been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date of implant: exact date of implant is unknown however occurred in (b)(6) 2017.Medical product: unknown femoral component: catalog#ni, lot#ni; unknown tibial tray: catalog#ni, lot#ni.G2: foreign: germany.Diligence is in process to determine whether the product is available for evaluation.The investigation is in progress.Upon receipt of additional information or completion of the investigation, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information at time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Upon receipt of additional information or completion of the investigation, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections have been updated: a1; b3; b4; b5; b7; d6; d9; d10; g3; h2; h6; h11.D10: medical product: unknown vanguard cr femoral component: catalog#ni, lot#ni; unknown vanguard cr tibial tray: catalog#ni, lot#ni; unknown refobacin cement: catalog#ni, lot#ni.The product has been received by zimmer biomet and the investigation is in process.Upon completion of the investigation, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an initial total knee arthroplasty.Subsequently, six (6) years and eight (8) months post-implantation, the patient experienced swelling and instability and underwent revision surgery.During the procedure, the polyethylene bearing was noted to have posterior wear.After extensive soft tissue balancing, the bearing was exchanged without reported complication.All available information has been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the proximal and distal surface to exhibit wear and damage.The device was submitted for further analysis.Analysis determined the damage to the bearing surface is potentially due to overloading, possibly exacerbated by a misalignment causing extensive overloading on the right-hand posterior condyle.Device history record was reviewed and no discrepancies related to the reported event were found.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: revision surgical notes: eua positive for instability, posterior wear noted on the inlay, new anterior stabilizing poly placed.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: left total knee arthroplasty with medial compartment narrowing and metal on metal appearance suggesting polyethylene wear.Root cause was unable to be determined.Complaint sample was evaluated and the reported event was confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information at time of this report.
|
| |
|