AVANOS MEDICAL INC. BALLARD CLOSED SUCTION SYSTEM FOR ADULTS, DOUBLE SWIVEL ELBOW; VAP CLOSED SUCTION CATHETERS & ACCESSORIES
|
Back to Search Results |
|
| Model Number |
221036-5 |
| Medical Device Problem Code |
Decoupling (1145)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported, the respiratory therapist performed a routine change of the closed suction system on a tracheostomized patient, at the moment of connecting the vial for instillation in its corresponding port, it detached and uncoupled until the internal spring was observed.It was immediately changed, ¿there was no damage or clinical or hemodynamic alteration to the patient." the device was reportedly in place for less than 20 hours.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 09 feb 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.H3 other text : device not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record for lot 20100335 was reviewed and the product was produced according to product specifications.The actual complaint product was not returned for evaluation; however, the reporter provided an image of the alleged device.Based on review of the provided image, the reported event could be confirmed, the root cause is traced to manufacturing.All information reasonably known as of 07 jun 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|