| Model Number |
5626 |
| Medical Device Problem Code |
Failure to Disconnect (2541)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a replacement procedure due to normal elective replacement indicator (eri), the set screw was unable to be loosened resulting in the lead being unable to disconnect from the device.The lead was capped and replaced to resolve the event.The patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of the setscrew could not be loosened to remove the lead was confirmed.Analysis revealed that calcified blood was filling the setscrew inset.The calcified blood was preventing the wrench from entering and engaging the setscrew inset to untighten the setscrew.Wrenches cannot penetrate the calcified blood.With calcified blood filling the inset all the wrench will do is strip the inset as in this case.For lab testing the calcified blood was removed out of the setscrew inset.A wrench then could be fully inserted into the inset.The setscrew could be untightened normally, and the stuck lead removed out of the connector.The blood came through a hole punched through the septum by the user of the torque wrench at time of implant.
|
| |
|