| Catalog Number |
705297 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "the instrument could not resist the force to reduce the bone and broke.".
|
| |
|
Additional Manufacturer Narrative
|
|
Please note correction to d4 (lot/serial no.).The reported event could be confirmed since the device was returned and matches the alleged failure.The device inspection revealed the following: the received reduction clamp was visually checked, and one of the tip of reduction clamp found completely broken off.Also, severe rubbing marks of usage clearly visible in the proximal areas of tip.The breakage of the tip shows the characteristics of a brittle failure with smooth surface and without any plastic deformation suggesting that the portion being subjected to excess stress leading to sudden breakage of the component.Furthermore, a hardness test according to rockwell on the returned item indicates the material being in accordance with the values in the material specification.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to be a user related issue as the breakage was caused due to the application of excess stress leading to sudden breakage.If more information is provided, the case will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported: "the instrument could not resist the force to reduce the bone and broke.".
|
| |
|
Search Alerts/Recalls
|