| Model Number |
A0393 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while performing the transurethral resection of the prostate procedure, when the high-frequency monopolar cable's cautery was activated, there was a spark, and the cord broke apart with a puff of smoke.The procedure was completed with a similar device.The device issue did not impact the outcome of the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
To date, the device has not been returned.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 5 years since the subject device was manufactured.Based on the results of the investigation, it is likely that the cable was used for longer than the specified 12-month timeframe.The cause of the reported issue is most likely due to wear and tear, in addition to improper device handling.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿a suitable replacement device must be provided during an application.¿ this supplemental report includes a correction to d9 to provide information that was inadvertently not included in the initial medwatch.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|