| Catalog Number |
72205110 |
| Medical Device Problem Code |
Flaked (1246)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a knee arthroscopy, the surgeon noticed a loose metal fleck on the side of the synovator platinum blade.The blade was pulled out from the knee joint and the metal fleck was wiped off.No piece fell inside the joint, since the surgeon noticed the problem in time.The procedure was completed using the same reported device.There was a surgical delay less than 30 minutes and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.However an image evaluation was performed and found two arthroscopic pictures of the side of the synovator platinum blade, where a unidentified material can be seen.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause of the reported event could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include debris from a concomitant device.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|