| Catalog Number |
320-15-01 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Aseptic loosening (4883)
|
| Date of Event |
08/27/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported by clinical shoulder study, the patient had a right tsa on (b)(6) 2020.The patient presented to the clinic on (b)(6) 2020 with increased pain after being overextended in physical therapy.Ct scan showed glenoid loosening.The patient was revised on (b)(6) 2020.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: pending investigation.D10: 6313744: 300-01-14 - equinoxe humeral stem primary press fit 14mm, 6513571: 320-10-00 - equinoxe reverse tray adapter plate tray +0 , 6509206: 320-02-38 - rs expanded glenosphere 38mm, +4mm offset , 6516750: 320-38-00 - equinoxe reverse 38mm humeral liner +0.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b3, d6b, h6, and h11.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The revision reported may have been due to reverse glenoid loosening.However, the reported glenoid loosening could not be confirmed.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|