EXACTECH, INC. EQUINOXE HUMERAL LINER, 40MM, +0; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Model Number |
HUMERAL LINER, 40MM, +0 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
02/17/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported via a clinical study, ¿equinoxe shoulder study¿, that a 55 yo female patient, initial left shoulder implanted on (b)(6) 2023, underwent a revision procedure on (b)(6) 2023.The patient dislocated their left shoulder anteriorly/inferiorly on (b)(6) 2023, with no injury reported, and underwent revision due to an acute inferior dislocation.The humeral liner was revised.The study indicates the event was definitely not related to the device or the procedure.The event was resolved on (b)(6) 2023.No further details.
|
| |
|
Additional Manufacturer Narrative
|
|
D10: concomitants:5.6969622 300-01-07 - equinoxe, humeral stem primary, press fit 7mm.A280070 320-31-40 - glenosphere, 40mm.A314350 320-10-00 - equinoxe reverse tray adapter plate tray +0.A316194 320-35-01 - small glenoid plate.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/technique.The reason for the reported revision due to dislocation cannot be conclusively determined; however, it may be due to soft tissue imbalance, rotator cuff failure, patient anatomy, implant positioning, and/or implant selection.Dislocation is a known risk, as outlined in the ifu.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|