Analysis of historical records orthofix checked the internal records related to the controls made on the device code 99-5415032 lot b1670879 before the market release.No anomalies have been found.The original lot, manufactured in 2021, was comprised of (b)(4) devices.40 of them have already been distributed to the market.According to orthofix historical records, no other notifications have been received from this specific device code.(mfr report 9680825-2024-00006).Orthofix checked the internal records related to the controls made on the device code 99-5415065 lot b1771062 before the market release.No anomalies have been found.The original lot, manufactured in 2021, was comprised of 30 devices.21 of them have already been distributed to the market.According to orthofix historical records, no other notifications have been received from this specific device code.(mfr report 9680825-2024-00007 and 9680825-2024-00008).Orthofix checked the internal records related to the controls made on the device code 99-5415070 lot b4301328 before the market release.No anomalies have been found.The original lot, manufactured in 2023, was comprised of (b)(4) devices.8 of them have already been distributed to the market.According to orthofix historical records, no other notifications have been received from this specific device code.(mfr report 9680825-2024-00009).Orthofix checked the internal records related to the controls made on the device code 99-5404532 lot b1771063 before the market release.No anomalies have been found.The original lot, manufactured in 2021, was comprised of (b)(4) devices.18 of them have already been distributed to the market.According to orthofix historical records, no other notifications have been received from this specific device code.(mfr report 9680825-2024-00010).Technical evaluation the technical evaluation on the returned devices is currently on going.As soon as further information and/or the results of the investigation are available, orthofix srl will provide you with a follow up report.Orthofix continues monitoring the devices on the market.
|
|
Analysis of historical records: orthofix checked the internal records related to the controls made on the device code 99-5415032 lot b1670879 before the market release.No anomalies have been found.The original lot, manufactured in 2021, was comprised of 62 devices.40 of them have already been distributed to the market.According to orthofix historical records, no other notifications have been received from this specific device code.(mfr report 9680825-2024-00006).Orthofix checked the internal records related to the controls made on the device code 99-5415065 lot b1771062 before the market release.No anomalies have been found.The original lot, manufactured in 2021, was comprised of 30 devices.21 of them have already been distributed to the market.According to orthofix historical records, no other notifications have been received from this specific device code.(mfr report 9680825-2024-00007 and 9680825-2024-00008).Orthofix checked the internal records related to the controls made on the device code 99-5415070 lot b4301328 before the market release.No anomalies have been found.The original lot, manufactured in 2023, was comprised of 20 devices.8 of them have already been distributed to the market.According to orthofix historical records, no other notifications have been received from this specific device code.(mfr report 9680825-2024-00009).Orthofix checked the internal records related to the controls made on the device code 99-5404532 lot b1771063 before the market release.No anomalies have been found.The original lot, manufactured in 2021, was comprised of 25 devices.18 of them have already been distributed to the market.According to orthofix historical records, no other notifications have been received from this specific device code.(mfr report 9680825-2024-00010).Technical evaluation: the returned devices, received on february 1, 2024, were examined by orthofix quality operations department.All screws were subjected to visual and functional check as per orthofix specification.In the visual check no issues were detected, except for signs of use, as scratches, residues, wear of threads.The dimensional check did not evidence any anomalies.Conclusions: in the visual check of the returned devices no issues were detected, except for signs of use, as scratches, residues, wear of threads.The dimensional check did not evidence any anomalies.The evidence collected suggest that the returned devices were originally conforming to orthofix specifications.No product problem was evidenced.Orthofix continues monitoring the devices on the market.
|