| Catalog Number |
620051180 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the humeral head trials broke on the inside where adapter trial fits.The version indicator plastic part cracked, sizer and glenoid trials cracked, inserter tip threads are worn.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary : no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Event or Problem Description
|
|
Additional information received: a.Was/were there any adverse consequence/s that affected the patient because of the reported event? -no adverse consequences.B.Did the event occur during the surgery? if yes, was there a surgical delay? what is the duration of the delay? -no surgical delay.C.Were there pieces broken off from the instrument? if yes, did it break into 2 or more pieces? - no pieces broken off.D.What part of the instruments were broken? -humeral head trials broke on the inside where adapter trial fits, version indicator plastic part cracked, sizer and glenoid trials cracked, inserter tip threads worn.E.Were all pieces of the broken instrument retrieved from the patient? -all pieces retrieved.
|
| |
|
Search Alerts/Recalls
|