ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0701-NTW
Medical Device Problem Code Incomplete Coaptation (2507)
Health Effect - Clinical Codes Mitral Valve Insufficiency/ Regurgitation (4451); Unspecified Tissue Injury (4559)
Date of Event 01/29/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information, the reported slda was due to patient condition.The reported recurrent mr was a cascading event of the reported slda.The cause of the reported tissue injury was unable to be determined.The reported patient effects of mitral regurgitation and tissue injury, as listed in the mitraclip system instructions for use, are known possible complications associated with mitraclip procedures.The reported hospitalization and unexpected medical intervention were results of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
It was reported that on may 16, 2022, a mitraclip procedure was performed to treat mixed mitral regurgitation (mr) with a grade of 4.A mitraclip ntw was implanted without issues, reducing the mr to a grade of 2.On (b)(6) 2024, an echocardiogram was performed and revealed recurrent mr with a grade of 4, a new tiny flail, and that the clip had detached from the posterior leaflet and remained attached to the anterior leaflet (single leaflet device attachment/slda).To stabilize the slda clip, another mitraclip was implanted, reducing the mr to a grade of 1-2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Common Device Name
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18696459
Report Number2135147-2024-00641
Device Sequence Number8508924
Product Code NKM
UDI-Device Identifier08717648230967
UDI-Public08717648230967
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/14/2022
Device Catalogue NumberCDS0701-NTW
Device Lot Number11014R110
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/29/2024
Initial Report FDA Received Date02/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR