ERIKA DE REYNOSA, S.A. DE C.V. STAY SAFE/SAFE LOCK CATHETER EXT. 12 IN.; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
| Catalog Number |
050-95001 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Codes |
Abdominal Pain (1685); Peritonitis (2252)
|
| Date of Event |
10/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On 24/jan/2024, a nx-stage area manager reported to post market surveillance that there was an allegation of a possible product deficiency involving the fresenius stay safe/safe lock catheter extension (product number 050-95001) in the (b)(6) clinic.The area manager provided the nurse contact email and additional follow-up was performed.On 25/jan/2024, the nurse contact reported that the (b)(6) clinic had a few patients that experienced a crack in the fresenius stay safe/safe lock catheter extension set (where the red end meets the copak) and as a result some of the patient¿s developed peritonitis.In an additional follow-up call to the nurse contact, it was reported that there were five peritoneal dialysis (pd)patients who developed peritonitis in association with the cracked pd catheter extension sets.The five peritonitis events occurred over time with the first event occurring on (b)(6) 2023 and the last reported event on (b)(6) 2024.Per clinic protocol, during a pd clinic visit each patient has the fresenius stay safe/safe lock catheter extension changed.Per the nurse, each crack in the extension set was discovered when the patient came into the pd clinic.Per the nurse, all the actual samples of the product involved in each instance of peritonitis have been discarded.Additionally, the nurse was not able to provide the lot#/#¿s for the involved products.There were no other allegations of additional adverse events related to the cracked extension sets.This clinic investigation was performed the capture the nature of the peritonitis event for this male patient with the initials (b)(6) per the nurse, the patient had the pd catheter extension set placed on (b)(6) 2023 (lot # unknown).On (b)(6) 2023, this patient was seen in the outpatient pd clinic for abdominal pain and cloudy effluent.The patient had a pd culture obtained (the same day) which grew corynebacterium species.The patient was initiated on a 3-week course of antibiotic therapy using vancomycin and fortaz (per clinic protocol) on an outpatient basis.During the pd clinic visit, a crack in the pd catheter extension set was discovered and peritonitis was attributed to the product.The patient had the pd catheter extension set changed and the patient continued pd therapy with the same cycler.The nurse confirmed the patient recovered from the peritonitis event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, a nx-stage area manager reported to post market surveillance that there was an allegation of a possible product deficiency involving the fresenius stay safe/safe lock catheter extension (product number 050-95001) in the (b)(6) clinic.The area manager provided the nurse contact email and additional follow-up was performed.On (b)(6) 2024, the nurse contact reported that the (b)(6) clinic had a few patients that experienced a crack in the fresenius stay safe/safe lock catheter extension set (where the red end meets the copak) and as a result some of the patient¿s developed peritonitis.In an additional follow-up call to the nurse contact, it was reported that there were five peritoneal dialysis (pd)patients who developed peritonitis in association with the cracked pd catheter extension sets.The five peritonitis events occurred over time with the first event occurring on 2/feb/2023 and the last reported event on 4/jan/2024.Per clinic protocol, during a pd clinic visit each patient has the fresenius stay safe/safe lock catheter extension changed.Per the nurse, each crack in the extension set was discovered when the patient came into the pd clinic.Per the nurse, all the actual samples of the product involved in each instance of peritonitis have been discarded.Additionally, the nurse was not able to provide the lot's for the involved products.There were no other allegations of additional adverse events related to the cracked extension sets.This clinic investigation was performed the capture the nature of the peritonitis event for this male patient with the initials (b)(6) per the nurse, the patient had the pd catheter extension set placed on (b)(6) 2023 (lot # unknown).On (b)(6) 2023, this patient was seen in the outpatient pd clinic for abdominal pain and cloudy effluent.The patient had a pd culture obtained (the same day) which grew corynebacterium species.The patient was initiated on a 3-week course of antibiotic therapy using vancomycin and fortaz (per clinic protocol) on an outpatient basis.During the pd clinic visit, a crack in the pd catheter extension set was discovered and peritonitis was attributed to the product.The patient had the pd catheter extension set changed and the patient continued pd therapy with the same cycler.The nurse confirmed the patient recovered from the peritonitis event.Additional follow up was conducted per manufacturing request to inquire the associated clinic¿s method (as well as chemical used) for cleaning the stay safe extension set while in use.The pdrn indicated that the blue end of the stay safe extension set (the port which connects to the fresenius cycler set or a manual pd bag for a pd treatment) is cleaned with alcavis for two minutes upon connection only.The nurse stated the patient is educated, trained, and performs this procedure.Per the nurse, the red end of the stay safe extension set (the end which connected to the patient¿s pd catheter) is cleaned daily with soap and water by the patient.Additionally, the red end of the stay safe extension set is cleaned with betadine when it is initially connected to the pd catheter copak adaptor.The nurse reiterated that the observed cracks occurred at the red end of the stay safe extension set.
|
| |
|
Search Alerts/Recalls
|
|
|