• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NOAH MEDICAL CORP. GALAXY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NOAH MEDICAL CORP. GALAXY SYSTEM Back to Search Results
Catalog Number GALB-001
Medical Device Problem Codes Entrapment of Device (1212); Difficult to Insert (1316); Difficult to Remove (1528)
Health Effect - Clinical Codes Cardiac Arrest (1762); Hemorrhage/Blood Loss/Bleeding (1888); Pneumonia (2011); Pneumothorax (2012); Septic Shock (2068)
Date of Event 02/21/2024
Type of Reportable Event Death
Additional Manufacturer Narrative
On (b)(6) 2024, during a noah galaxy assisted bronchoscopy procedure a patient experienced bleeding, cardiac arrest, and bilateral pneumothoraces.The incident prompted a thorough investigation to understand the root causes and contributing factors.Upon reviewing the information gathered from the event and analyzing the video log, it was observed that the physician encountered challenges in navigating a difficult area within the patient's upper left lobe(lul).The physician was unable to fully reach the intended lesion location with the bronchoscope contributing to scope buckling.Throughout the procedure, there were six instances of scope buckling error notifications.Ultimately, the repeated attempts by the physician to drive the scope forward and continued acknowledgement of the scope buckling workflow inhibitor error message resulted in the creation of a kink in the bronchoscope shaft exhibiting a fishhook shape, further complicating the procedure.Extraction of the bronchoscope in the buckled and fishhook shape proved challenging, requiring the physician to manually release the bronchoscope from the robot and push and rotate until the bronchoscope tip was able to pass back through the endotracheal tube and be removed from the patient.The results of the investigation determined that the adverse event occurred due to a use error where multiple (six) acknowledgments of scope buckling error messages occurred, but the user failed to follow the on-screen and ifu instructions to utilize fluoroscopic imaging to check the shape of the scope.Each time the system error message occurred, the use of fluoroscopy was not utilized to check the condition of the bronchoscope or the cause of the scope buckling, the user acknowledged the workflow inhibitor message and proceeded forward with the case without following the on-screen and ifu instructions related to scope buckling warnings.Based on the investigation, scope buckling was not a result of device malfunction or defective scope.It was the repeated use error of dismissing the error message without performing the recommended fluoroscopy to verify the scope position that caused damage to the scope and ultimately contributed to harming the patient.Specifically, the galaxy ifu (10000752 rev.F) contains the following warning: "do not continue inserting the scope if scope buckling is detected.Continued insertion may result in patient injury." the on-screen error message (which is also included in the ifu alerts table) contains the following verbiage: "scope buckling detected.Please retract until scope insertion is responsive to controls, then try again.If this problem persists, please take a fluoro image to check the shape of the scope." a comprehensive investigation also included a review of the lot history record associated with the medical device used in the procedure.Notably, no deviations or non-conformances were identified for this lot.The device was manufactured and shipped in strict adherence to specifications.Because of the complications arising from the procedure, the patient is currently hospitalized.The medical team is actively monitoring and addressing the patient's condition.Patient's condition will continue to be monitored until release from the hospital and will be documented in the complaint record.
 
Event or Problem Description
On (b)(6) 2024, during a bronchoscopy procedure assisted by the noah galaxy system, a patient experienced bleeding, cardiac arrest, and bilateral pneumothoraces.The target lesion was located in the right upper lobe (rul).The physician attempted to navigate to the lesion using the noah medical galaxy system, the procedure was impeded due to scope buckling, resulting in a kinked scope that proved challenging to remove from the patient's body.Consequently, the patient suffered adverse events.Presently, the patient remains hospitalized as a result of complications arising from the procedure.The medical team is actively monitoring and addressing the patient's condition.For additional details, please refer to manufacturer's narrative for investigation and rootcause.
 
Event or Problem Description
This supplemental report serves as a follow- up to an event documented under mdr 3021325287-2024-00001.At the time of the initial mdr submission, the patient was under intensive care unit (icu) supervision and being monitored by a physician.She recovered enough to be moved from icu to step down.Subsequently the patient passed away due to medical complications arising from septic shock secondary to pneumonia and respiratory failure.Noah medical became aware of this outcome during the follow-up on the patient's heath condition.Additionally, the attending physician did not attribute the patient's death to any usage or malfunction of the noah galaxy device, but rather to her pre-existing heath conditions.Galaxy system was used in procedure and may have contributed to the initial injury.The initial submission of the supplemental report was made on 11march2024.Support from emdr helpdesk requested resubmission.Resubmitting the report on november 21, 2024.
 
Additional Manufacturer Narrative
This supplemental report serves as a follow-up to an event documented under mdr 3021325287-2004-00001.At the time of the initial mdr submission, the patient was under intensive care unit (icu) supervision and being monitored by a physician.She recovered enough to be moved from icu to step down.Subsequently the patient passed away due to medical complications arising from septic shock secondary to pneumonia and respiratory failure.Noah medical became aware of this outcome on 21feb2024 during the follow-up on the patient's health condition.Importantly, the attending physician did not attribute the patient's death to any usage or malfunction of the noah galaxy device, but rather to her pre-existing health conditions.Galaxy system was used in procedure and may have contributed to the initial injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAXY SYSTEM
Common Device Name
GALAXY SYSTEM
Manufacturer (Section D)
NOAH MEDICAL CORP.
2075 zanker road
san jose CA 95131
Manufacturer (Section G)
NOAH MEDICAL CORP.
2075 zanker road
san jose CA 95131
Manufacturer Contact
patricia apps
2075 zanker road
san jose, CA 94070
3412266514
MDR Report Key18699853
Report Number3021325287-2024-00001
Device Sequence Number6143693
Product Code EOQ
UDI-Device Identifier00850048825000
UDI-Public01008500488250001020311070317240500
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K223144
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/07/2024
Device Catalogue NumberGALB-001
Device Lot Number2023110703
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/19/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer02/21/2024
Initial Report FDA Received Date02/13/2024
Supplement Report FDA Received Date11/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/07/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Hospitalization; Required Intervention; Other;
Patient Age76 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-