The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd drive pcb.In addition, our technician confirmed that the control body corroded, the ccd drive pcb corroded, the electrical pin connector corroded, the lg connector corroded, the remote control buttons leak, the remote control buttons cut, and the electrical pin connector defective; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|