The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd module with drive pcb.In addition, our technician confirmed that the lg connector fluid damage, the light guide cable buckled, the ccd drive pcb corroded, the deflector washing socket deformed, the operation channel (primary) leak, the operation channel (primary) resistance, and the suction channel kink; however, these defects are not the main cause, and or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and or the risk analysis results, it was evaluated to submit mdr.
|