Model Number MID-C 125
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Production process analysis: a review of the device history record confirmed that the device was manufactured and tested according to relevant procedures, and shipped according to manufacturer's specifications.User (surgeon and patient) information analysis: patient (b)(6) index procedure was performed on (b)(6) 2022.Patient (b)(6) underwent revision surgery on (b)(6) 2023 due to the initial device reaching maximum elongation; device was replaced with a longer device.(ref: complaint (b)(4) during monitoring, as part of the pas study, apifix identified on (b)(6) 2024 that patient (b)(6) was scheduled for a removal surgery on (b)(6) 2024 due to screw pull-out and subsequent implant migration.According to the report, 'the most distal of the proximal screws has plowed through the pedicle causing increase in the kite angle'.On (b)(6) 2024 apifix identified that patient (b)(6) underwent removal surgery on (b)(6) 2024.No report of patient harm/complications was received.Risk assessment: reoperation events are a known risk that was assessed and recorded by the product risk management file.The risk of screw pullout has been assessed and found to be acceptable the current screw pullout rate due to any reason is in line with the rate reported in the literature for this type of complication as described in the company's cer (clinical evaluation report).The risk has been quantified, characterized, and documented as acceptable within a full risk assessment.The event of screw pull-out is addressed in the ifu as potential risks associated with the mid-c system.The explanted device is expected to be returned to the manufacturer and will be evaluated; upon its completion, once new information comes to light, then a supplemental medwatch report will be submitted.
 
Event or Problem Description
During monitoring, as part of the pas study, apifix identified on (b)(6) 2024 that patient (b)(6) (pas (b)(6) was scheduled for a removal surgery on (b)(6) 2024 due to screw pull-out and subsequent implant migration.According to the report, 'the most distal of the proximal screws has plowed through the pedicle causing increase in the kite angle'.
 
Additional Manufacturer Narrative
Return analysis: the explanted device was returned and was subjected to cleaning, steam sterilizing, and engineering evaluation.During inspection, the device appeared normal with no obvious signs of failure.The extender was removed, and the spherical rings were inspected for wear.No visible wear was found.A marking was found on the implant, which is suspected to be from the cleaning process implant and not indicative of any device failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MID-C 125
Common Device Name
POSTERIOR RATCHETING ROD SYSTEM
Manufacturer (Section D)
APIFIX LTD.
1 hacarmel st.
kochav yokneam bldg
yokneam ellit, 20692 07
IS  2069207
Manufacturer (Section G)
APIFIX LTD.
1 hacarmel street
kochav yokneam bldg
yokneam elit, 20692 07
IS   2069207
Manufacturer Contact
alan vaisman
1 hacarmel st.
kochav yokneam bldg
yokneam ellit, 20692-07
IS   2069207
MDR Report Key18701639
Report Number3013461531-2024-00010
Device Sequence Number20369452
Product Code QGP
UDI-Device Identifier07290018128046
UDI-Public07290018128046
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
H170001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMID-C 125
Device Catalogue NumberMUS-125-050
Device Lot NumberAF-03-080-21
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/14/2024
Supplement Report FDA Received Date08/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age12 YR
Patient SexFemale
-
-