| Model Number |
SN60WF |
| Medical Device Problem Code |
Failure to Unfold or Unwrap (1669)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported with a description of damaged intraocular lens (iol).Additional information has been received and stated that the lens did not open correctly in the eye and procedure was completed with new lens during initial procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.Associated products were not provided.It is unknown if qualified products were used.The product investigation could not identify a root cause for the reported complaint.The product has not been received to evaluate.Associated products were not provided.It is unknown if qualified products were used.The instruction for use (ifu) instructs: company foldable intra ocular lens (iol) are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.Follow up attempts were made.The account did not provide any further information at this time.File will be reopened when new information or the product is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|