On 30th january, 2024 getinge became aware of an issue with one of surgical lights - volista access 600.As it was stated, two sides covers and a tab were missing.The designated complaint unit employee confirmed based on photographic evidence the two sides covers and the dust covers were not present on the device and have some missing particles.There was no injury reported, however, we decided to report the issue in abundance of caution as any parts or particles falling off into sterile field or during procedure may cause contamination in case of reoccurrence.
|
|
The correction of d1 brand name, d4 version of model #, d4 catalog #, d4 unique identifier (udi) #, h3a device evaluated by manufacturer, h3b device not eval provide code, h3c if other provide code - explain fields deems required.This is based on the internal evaluation.Previous d1 brand name: volista access corrected d1 brand name: volista previous d4 version of model # ard568806905 corrected d4 version of model # ardvcs209029a previous d4 catalog # ard568806905 corrected d4 catalog # none previous d4 unique identifier (udi) #: n/a corrected d4 unique identifier (udi) #: (b)(6).Previous h3a device evaluated by manufacturer: no corrected h3a device evaluated by manufacturer: yes previous h3b device not eval provide code: other corrected h3b device not eval provide code: n/a previous h3c if other provide code - explain: device not returned to manufacturer corrected h3c if other provide code - explain: n/a.Getinge became aware of an issue with one of surgical lights - volista access ii 600.As it was stated, two sides covers and a tab were missing.The designated complaint unit employee confirmed based on photographic evidence the two sides covers and the dust covers were not present on the device and have some missing particles.There was no injury reported, however, we decided to report the issue in abundance of caution as any parts or particles falling off into sterile field or during procedure may cause contamination in case of reoccurrence.Based on the information collected, it was established that when the event occurred, the surgical light did not meet its specification and in this way the device contributed to event.Provided information does not indicate if upon the event occurrence, the device was or was not being used for patient treatment or diagnosis.A root cause analysis for investigated problem was performed by subject matter experts and concluded that: according to the photographic evidence the covers involved had broken tabs and could not be assembled, that's why they were missing.The spring arm covers involved have probably been broken as a result of a collision or incorrect assembly or disassembly.The manufacturer recommends following instructions in the volista installation manual 01814, which describes how to fit the covers on the spring arm.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
|