On 19-jan-2024, abbott point of care (apoc) was contacted by a customer regarding i-stat b-hcg cartridges that yielded a positive result on a 43 year old female patient with vomiting, emesis, diarrhea, and abdominal pain.There was no additional patient information at the time of this report.Return product is not available for investigation.Method; date; collected; tested; result (iu/ml) i-stat, (b)(6) 2023, 10:59am, 11:10am & 35.2.Roche, (b)(6) 2023, ni, ni & <5.0.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.I-stat positive cut-off values: b-hcg < 5.0 iu/l, negative.5.0 < ss-hcg < 25.0 iu/l, indeterminate.B-hcg >25.0 iu/l, positive.Intended use: the i-stat® total beta-human chorionic gonadotropin (b-hcg) test is an in vitro diagnostic test for the quantitative and qualitative determination of b-hcg in venous whole blood or plasma samples using the i-stat 1 analyzer systems.The test is intended to be used as an aid in the early detection of pregnancy and is for prescription use only.
|