| Model Number |
7590 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the severely calcified left anterior descending artery.A 2.50mm x 20mm maverick balloon catheter was advanced for dilatation.However, during inflation, the balloon burst.The procedure was completed with another of the same device.There were no patient injuries reported, and the patient condition was good post-procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the severely calcified left anterior descending artery.A 2.50mm x 20mm maverick balloon catheter was advanced for dilatation.However, during inflation, the balloon burst.The procedure was completed with another of the same device.There were no patient injuries reported, and the patient condition was good post-procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a maverick 2mr balloon catheter.Visual and microscopic inspection revealed no damage or irregularity.There was contrast present within the inflation lumen, blood in the guidewire lumen, and the balloon was loosely folded.The device was prepped with an inflation device filled with water and connected to the calibrated pressure gage to inflate the device.The device inflated to rated burst pressure and deflated with no issue.
|
| |
|