Model Number X-SPM-03
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 12/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare professional reported that a residual error with post operative treatment results was less than one diopter, the axis offset by ninety degrees (out of range) in the left eye of a patient, after surgery.After several attempts to align the eye and was often out of range.
 
Additional Manufacturer Narrative
H.3., h.6.: investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The manufacturer's product subject matter expert concluded that the surgeon had set an incorrect microscope orientation leading to the axis being displayed at a different position for the first surgery.However, during the touch up surgery, surgeon has corrected this and selected the correct microscope orientation and realigned the lens.The company representative is aware of the findings and will inform the findings and retrain the customer on the importance of selecting the correct orientation.No sample is expected to be returned for evaluation.The investigation is therefore concluded based on the provided information.The root cause is an incorrect microscope orientation selected at the start of surgery, causing the axis to be displayed at a different position (i.E., user handling), system met specifications.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERION DIGITAL MARKER M
Common Device Name
MARKER, OCULAR
Manufacturer (Section D)
WAVELIGHT GMBH (AGPS)
rheinstrasse 8
teltow 14513
GM  14513
Manufacturer (Section G)
WAVELIGHT GMBH (AGPS)
rheinstrasse 8
teltow 14513
GM   14513
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18703617
Report Number3010300699-2024-00004
Device Sequence Number7217577
Product Code HLG
UDI-Device Identifier00380659982705
UDI-Public00380659982705
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 07/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberX-SPM-03
Device Catalogue Number8065998270
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/26/2024
Supplement Date Received by Manufacturer06/11/2024
Initial Report FDA Received Date02/14/2024
Supplement Report FDA Received Date07/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/24/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age35 YR
Patient SexFemale
-
-