Model Number AR-TL08E
Medical Device Problem Codes Corroded (1131); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned and the evaluation is ongoing.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if additional information is provided by the user facility.
 
Event or Problem Description
The customer reported the subject device had water invasion and poor image quality.No patient harm reported.
 
Event or Problem Description
The customer reported the subject device had water invasion and poor image quality.No patient harm reported.
 
Additional Manufacturer Narrative
The device was returned to olympus for evaluation and the reported allegation was confirmed.The device evaluation found operation failure, lens flooding, the scope mount sticking due to corrosion, and body corrosion.Based on the results of the investigation, a definitive root cause cannot be identified.A device history review revealed no issues that could have caused or contributed to the reported issue.This issue is addressed in the instructions for use (ifu): ¦preparation and inspection ¿ be sure to perform the following inspections and discontinue use if any abnormality is found.Continue to use the product under such conditions may cause the device to malfunction during use.- turn the scope lock and make sure it turns smoothly.- pinch the scope lock and lock ring knob and rotate the lock ring to make sure there is no snagging.[section where ifu applicable for phenomenon 2] ¦*important basic precautions ¿2.Do not use excessive force when using this device.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEO ADAPTER
Common Device Name
VIDEO ADAPTER
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18704238
Report Number3002808148-2024-30725
Device Sequence Number15291846
Product Code GCJ
UDI-Device Identifier04953170023972
UDI-Public04953170023972
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K955404
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAR-TL08E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/17/2024
Supplement Date Received by Manufacturer02/26/2024
Initial Report FDA Received Date02/14/2024
Supplement Report FDA Received Date03/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/01/1950
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-