| Model Number |
AR-TL08E |
| Medical Device Problem Codes |
Corroded (1131); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned and the evaluation is ongoing.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
The customer reported the subject device had water invasion and poor image quality.No patient harm reported.
|
| |
|
Event or Problem Description
|
|
The customer reported the subject device had water invasion and poor image quality.No patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation and the reported allegation was confirmed.The device evaluation found operation failure, lens flooding, the scope mount sticking due to corrosion, and body corrosion.Based on the results of the investigation, a definitive root cause cannot be identified.A device history review revealed no issues that could have caused or contributed to the reported issue.This issue is addressed in the instructions for use (ifu): ¦preparation and inspection ¿ be sure to perform the following inspections and discontinue use if any abnormality is found.Continue to use the product under such conditions may cause the device to malfunction during use.- turn the scope lock and make sure it turns smoothly.- pinch the scope lock and lock ring knob and rotate the lock ring to make sure there is no snagging.[section where ifu applicable for phenomenon 2] ¦*important basic precautions ¿2.Do not use excessive force when using this device.Olympus will continue to monitor the field performance of this device.
|
| |
|
Search Alerts/Recalls
|