| Model Number |
VICM5 13.2 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Intraocular Pressure Increased (1937)
|
| Date of Event |
12/09/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - work order search: no similar complaint type events were reported for units within the same lot.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 13.2mm, vicm5 13.2, -10.0 diopter, implantable collamer lens was implanted into the patient's right eye (od).Elevated iop was observed.Lens remains implanted.Additional information has been requested but none has been forthcoming.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: the reporter indicated that the surgeon implanted a 13.2mm vicm5_13.2; -10.0 diopter implantable collamer lens into the patient's right eye (od) on (b)(6) 2023.There was excessive vaulting, pupil block with high iop, medical management performed and on (b)(6) 2024 the lens was exchanged with a shorter length lens and the problem was resolved.The cause of the event was reported as the device.Claim#: (b)(4).
|
| |
|