• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LARGE SCREWDRIVER HANDLE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LARGE SCREWDRIVER HANDLE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 71173547
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal reference number: (b)(4).H10: smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.
 
Event or Problem Description
It was reported that, during an internal fixation procedure, the handle part of one (1) large screwdriver handle snapped whilst screwing in the large frag screw during plating.It snapped between the metal and wooden part.As it only broke into two, the pieces were retrieved by hand.The procedure was resumed, after a non-significant delay, with a s+n back-up device.Patient was not harmed as a consequence of this problem.
 
Additional Manufacturer Narrative
Section h10: the associated device was returned and evaluated.The visual inspection revealed the connection point of the device broke off.This piece was returned.The device shows signs of significant wear and use.This inspection was conducted by naked eye.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LARGE SCREWDRIVER HANDLE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18705883
Report Number1020279-2024-00359
Device Sequence Number6349997
Product Code HSB
UDI-Device Identifier03596010519580
UDI-Public03596010519580
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71173547
Device Lot Number18KTMH139
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2024
Supplement Date Received by Manufacturer03/15/2024
Initial Report FDA Received Date02/14/2024
Supplement Report FDA Received Date03/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/12/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-