• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 16 CM; CATHETER PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 16 CM; CATHETER PERCUTANEOUS Back to Search Results
Catalog Number CS-12122-F
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported "one of swg was faulty and cannot used (bent), and then required a new set" the current status of the patient is "unknown" at ths time.The customer has not returned our request for additional information at this time.
 
Event or Problem Description
It was reported "one of swg was faulty and cannot used (bent), and then required a new set" the current status of the patient is "unknown" at ths time.The customer has not returned our request for additional information at this time.
 
Additional Manufacturer Narrative
(b)(4).The report that the spring wire guide was kinked was confirmed through examination of the returned sample.The customer returned one, opened hemodialysis kit including one swg within its advancer for analysis.The swg cap and straightener tube, components of the swg assembly, were not returned.Signs of use were observed on the returned sample, which according to additional information received from the customer, likely occurred in the clinical setting.Visual analysis confirmed that there was one kink towards the distal j-bend of the guide wire.The distal j-bend was not misshapen and was intact.Microscopic examination confirmed the defect.Both welds were present and were observed to be full and spherical.No other defects or anomalies were observed with the guide wire or tubing.The kink in the guide wire was located 565 mm via calibrated ruler from the proximal tip.The overall length of the guide wire measured 601 mm via calibrated ruler which was within the specification of 596-604 mm per guide wire product drawing.The outer diameter (od) of the guide wire measured 0.841 mm via calibrated micrometer which was within the specification of 0.838-0.877 mm per guide wire product drawing.Functional inspection was performed per ifu statement, "advance guidewire into arrow raulerson syringe approximately 10 cm until it passes through syringe valves or into introducer needle." a manual tug test confirmed both welds were intact.The instructions-for-use (ifu) provided with the kit warns the user, "do not use if package is damaged".A device history record review was performed, and no relevant findings were identified.During full assembly, the kink on the guide wire would have been located within the straightener tube/swg cap during packaging, shipping, and storage, protecting it from damage.However, the end cap and straightener tube were missing from the returned assembly.Based on the customer report and sample received, the potential root cannot be determined as it could not be confirmed when the damage to the guidewire occurred.Teleflex will continue to monitor and trend for complaints of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 16 CM
Common Device Name
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18707258
Report Number3006425876-2024-00116
Device Sequence Number7271609
Product Code DQY
UDI-Device Identifier00801902100108
UDI-Public00801902100108
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K862056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCS-12122-F
Device Lot Number71F23E0581
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2024
Supplement Date Received by Manufacturer03/14/2024
Initial Report FDA Received Date02/14/2024
Supplement Report FDA Received Date03/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED; NONE REPORTED
Patient SexUnknown
-
-