DENTSPLY LLC CAVITRON JET PLUS TAP-ON(DNA) G137; SCALER, ULTRASONIC Back to Search Results
Catalog Number G137
Medical Device Problem Codes Restricted Flow rate (1248); Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event or Problem Description
While using a cavitron jet plus g137, they allege that the inserts are getting hot with potential water flow issue, no injury resulted.
 
Additional Manufacturer Narrative
Evaluation tech: gpb unit serial number-((b)(6)) foot pedal serial number-((b)(6)) jet-mate/handpiece/sterimate lot number-#1-(2023080)-#2-(07130) handpiece cable lot number-(09200) root cause: w4o water sol,iso valve leaking continuous water leaking due to debris buildup in the water system not allowing the solenoid to close completely damaged water flow control on the handpiece cable (water flow turn knob is loose and debris build up is restricting water flow causing the tips to get hot/warm) air manifold not exhausting properly, also the air manifold has bad regulation resulting in inconsistent powder flow, leaking air from manifold two clogged duckbill filter causing poor air/powder flow flattened pinch tube restricting air/powder flow powder bowl needs to be cleaned and retubed foot control pad is worn and needs to be replaced water supply hose missing black sleeves to secure tubing to filter luers note: will replace damaged/worn components and recalibrate unit to factory specs upon estimate approval.For proper evaluation and testing please always send in all parts that go along with the unit to be looked at.Loaner fee included on estimate additional notes: not enough water was getting to the jet-mate from the debris building up in the handpiece cable and the water solenoid and furthermore the jet-mate was leaking which was making poor water flow connection to the insert that was making the jet-mate and insert heat up with not enough water to the insert and the jet-mate.Jet air polishing insert that was sent in by customer has been evaluated and tested.The jet air polishing insert is bent and causing poor air/water flow through the jet air polishing insert.This insert is not in working manufactured standards.Jet-mate lot code-2023080-good to use it is up to manufactured specs jet-mate lot code-07130-good to use it is up to manufactured specs items not received for testing and evaluation no auxiliary foot pedal charging cord.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON JET PLUS TAP-ON(DNA) G137
Common Device Name
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
york, PA 17401
7178494593
MDR Report Key18707451
Report Number2424472-2024-00010
Device Sequence Number15235521
Product Code ELC
UDI-Device IdentifierD00381614281
UDI-PublicD00381614281
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K052334
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberG137
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/14/2024
Supplement Date Received by Manufacturer02/14/2024
Initial Report FDA Received Date02/14/2024
Supplement Report FDA Received Date03/15/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-