| Catalog Number |
INC-11129-132 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3 other text : the device is not available to the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure the subject catheter would not advance and got broken/fractured.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.No further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the results of the device history record (dhr) review, there is no indication that the device, labeling or packaging failed to meet its specifications when released.Visual and functional inspections were not performed as the device was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.As the device was not returned, the reported events: ¿catheter shaft broken/fractured during use¿ and ¿catheter shaft friction¿ could not be confirmed.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure the subject catheter would not advance and got broken/fractured.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.No further information available.
|
| |
|