• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. SEEKER CROSSING SUPPORT; SUPPORT CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. SEEKER CROSSING SUPPORT; SUPPORT CATHETER Back to Search Results
Catalog Number SK15018
Medical Device Problem Codes Break (1069); Material Puncture/Hole (1504); Detachment of Device or Device Component (2907); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10 : as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 01/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that during a recanalization procedure in right femoral artery, the guidewire allegedly failed to pass through the support catheter.It was further reported that the guidewire allegedly poked out at the half-section of the support catheter.Reportedly, the support catheter was allegedly broken.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.Three photos were provided and reviewed.The first photo shows the seeker device within its packaging.The labeling matches with the information available in trackwise.The catheter shaft is seen completely detached in two segments.The second and third photos also show the catheter shaft of the device completely detached in two segments.Therefore, the investigation is unconfirmed for the reported break and material hole but confirmed for the identified detachment as the catheter shaft was seen completely detached.The investigation is also inconclusive for the reported device-device incompatibility as functional testing for the same could not be performed.A definitive root cause for the reported break, material hole, device-device incompatibility and identified detachment could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: b5, d4 (expiration date: 01/2025), g3, h6 (device, method).H11: h6 (result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that during a recanalization procedure in the right femoral artery via the retrograde approach, the guidewire allegedly failed to pass through the support catheter.It was further reported that the guidewire allegedly got poked out at the half-section of the support catheter.Reportedly, the support catheter was allegedly broken.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEEKER CROSSING SUPPORT
Common Device Name
SUPPORT CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe 85281
Manufacturer (Section G)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18708711
Report Number2020394-2024-00203
Device Sequence Number7205184
Product Code DQY
UDI-Device Identifier00801741067914
UDI-Public(01)00801741067914
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K103367
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSK15018
Device Lot NumberVTHN0413
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/24/2024
Supplement Date Received by Manufacturer04/25/2024
Initial Report FDA Received Date02/15/2024
Supplement Report FDA Received Date04/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-