• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5530-G-209
Medical Device Problem Codes Degraded (1153); Fracture (1260); Material Fragmentation (1261); Naturally Worn (2988)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 01/23/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.H3 other text : not returned.
 
Event or Problem Description
Patient had received a primary tka in 2012 though underwent revision surgery due to ongoing instability issues.Initial examination during surgery ((b)(6 2024) demonstrated increased medial/lateral laxity.Surgeon discovered fragments of initial cr size #2 9mm poly around the knee joint with significant wear on the posteromedial and posterolateral surfaces.Surgeon removed fragments of insert and replaced with a cs size #2 13mm poly.
 
Additional Manufacturer Narrative
Reported event: an event regarding wear & crack/fracture involving a triathlon insert was reported.The event for wear and crack/fracture were confirmed via evaluation of the returned device.Method & results: product evaluation and results: visual inspection of the returned device indicated that the damage present on the anterior portion of both condyles was consistent with delamination and material loss, two small pieces fractured off.The damage on the articulating surface of the insert is consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to increased medial/lateral laxity.Surgeon discovered fragments of poly around the knee joint with significant wear on the posteromedial and posterolateral surfaces.Visual inspection of the returned device indicated that the damage present on the anterior portion of both condyles was consistent with delamination and material loss, two small pieces fractured off.The damage on the articulating surface of the insert is consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Patient had received a primary tka in 2012 though underwent revision surgery due to ongoing instability issues.Initial examination during surgery (b)(6) 2024 demonstrated increased medial/lateral laxity.Surgeon discovered fragments of initial cr size #2 9mm poly around the knee joint with significant wear on the posteromedial and posterolateral surfaces.Surgeon removed fragments of insert and replaced with a cs size #2 13mm poly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18710375
Report Number0002249697-2024-00233
Device Sequence Number8276347
Product Code MBH
UDI-Device Identifier07613327046410
UDI-Public07613327046410
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K172634
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2016
Device Catalogue Number5530-G-209
Device Lot NumberLCQ915
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/23/2024
Supplement Date Received by Manufacturer05/28/2024
Initial Report FDA Received Date02/15/2024
Supplement Report FDA Received Date06/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/02/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age85 YR
Patient SexUnknown
-
-