• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. LOW PROFILE LOCKING SCREW 3.5 MM X 20 MM; BONE FIXATION SCREW/PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. LOW PROFILE LOCKING SCREW 3.5 MM X 20 MM; BONE FIXATION SCREW/PLATE Back to Search Results
Model Number LOW PROFILE LOCKING SCREW 3.5 MM X 20 MM
Medical Device Problem Codes Positioning Failure (1158); Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during a lapidus surgery the screw did not fit properly into the plate.The screw was inserted either way.There was no harm for patient, operator or third party reported.It was not necessary to switch the surgical technique or do a second surgery.Update 02-feb-2024 dw: further information were provided that while the surgeon inserted the second screw it broke off.The broken piece was retrieved and the surgery was finished successfully with an additional screw with the same part number.
 
Additional Manufacturer Narrative
Complaint is confirmed.Visual inspection identified the damage on the threads of the screw of the low profile locking screw, titanium, 3.5 x 20 mm, sterile, im had stripped/damaged.Functional testing was not performed due to the damage to the device.The most likely cause for the reported failure can be attributed to user error due to misalignment insertion and/or prying/leveraging the screw during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOW PROFILE LOCKING SCREW 3.5 MM X 20 MM
Common Device Name
BONE FIXATION SCREW/PLATE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key18710570
Report Number1220246-2024-00982
Device Sequence Number8192171
Product Code HWC
UDI-Device Identifier00888867050921
UDI-Public00888867050921
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K143614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 12/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLOW PROFILE LOCKING SCREW 3.5 MM X 20 MM
Device Catalogue NumberAR-8935L-20S
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2024
Supplement Date Received by Manufacturer06/25/2024
Initial Report FDA Received Date02/15/2024
Supplement Report FDA Received Date12/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-