| Catalog Number |
5510F502 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/25/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.H3 other text : device not returned to the manufacturer.
|
| |
|
Event or Problem Description
|
|
Primary procedure, right knee.Knee was planned for size 5 right pressfit cr femur.A size 5 right cemented cr femur was implanted.Surgeon was notified post-op.Patient had left or, was notified and returned for revision of the cemented femur for the pressfit femur the next day.
|
| |
|
Event or Problem Description
|
|
Primary procedure, right knee.Knee was planned for size 5 right pressfit cr femur.A size 5 right cemented cr femur was implanted.Surgeon was notified post-op.Patient had left or, was notified and returned for revision of the cemented femur for the pressfit femur the next day.Update: the patient was revised to a cemented femur with cement instead of a pressfit implant.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding incorrect selection involving a triathlon femoral component was reported.The event was confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that a cemented femoral component was implanted in the patient's knee instead of a press-fit (cementless) femoral component.The patient was revised the following day to correct the error.The provided implant sheet was reviewed and confirmed that the reported device, a size 5 triathlon cruciate retaining femoral component, right side, catalog: 5510-f-502, lot: clelu was implanted in the patient's right knee on january 25, 2024.The reported event is considered to be the result of user error as no manufacturing-related product problem was found given the information provided.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|