It was observed that during the device evaluation, the coated portion of the single use sphincterotome v (distal wireguided) cutting wire was torn, and the broken portion was scorched and melted from touching metal when activated.There were no reports of patient involvement.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event could not be determined although it can be presumed that the event occurred due to the coating of the wire came into contact with the distal end of the endoscope activating electric conduction causing the cutting wire to become hot and break.The event could have been detected/prevented by following the instructions for use: ¿since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.¿when inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.¿do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.¿if you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion.Olympus will continue to monitor field performance for this device.
|