On (b)(6) 2024 getinge became aware of an issue with one of surgical lights - volista standop.As it was stated, the missing protective caps were installed (dust covers).There is no information about the circumstances of the missing parts.There was no injury reported, however, we decided to report the issue in abundance of caution as any parts or particles falling off into sterile field or during procedure may cause contamination in case of reoccurrence.
|
|
The correction of d1 brand name, d4 version of model #, d4 catalog #, d4 unique identifier (udi) #, h3a device evaluated by manufacturer, h3b device not eval provide code, h3c if other provide code - explain fields deems required.This is based on the internal evaluation.Previous d1 brand name: volista standop; corrected d1 brand name: standop volista®; previous d4 version of model # ard568832960; corrected d4 version of model # ardvst229039a; previous d4 catalog # ard568832960 ; corrected d4 catalog # none; previous d4 unique identifier (udi) #: n/a; corrected d4 unique identifier (udi) #: (b)(4); previous h3a device evaluated by manufacturer: no; corrected h3a device evaluated by manufacturer: yes; previous h3b device not eval provide code: other; corrected h3b device not eval provide code: n/a; previous h3c if other provide code - explain: device not returned to manufacturer; corrected h3c if other provide code - explain: n/a.Getinge became aware of an issue with one of surgical lights - volista standop.As it was stated, the missing protective caps were installed (dust covers).There is no information about the circumstances of the missing parts.There was no injury reported, however, we decided to report the issue in abundance of caution as any parts or particles falling off into sterile field or during procedure may cause contamination in case of reoccurrence.Missing part was replaced (ard367827555-sat/pwd/pwdii/vst-ac2000 sf dustcovers ) and device returned to usage.Based on the information collected, it was established that when the event occurred, the surgical lights did not meet its specification, since cover detachment could be considered as technical deficiencies, and in this way the device contributed to event.It is unknown if claimed devices were or were not being used for patient treatment or diagnosis when the event took place.According to subject matter expert at manufacturer¿s, the most probable root causes are : non-conformity of the metal covers assembly.Degradation of the metal covers.Improper use (collision with another device).Maquet sas analysis shows that the metal strip comes out of the covers when it is not clipped properly.In the scope of our continuous improvement policy, maquet sas initiated a modification file ((b)(4)) to include this dust cover fitting procedure in the technical documentations with all spring arms.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
|