• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PT; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PT; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 2777
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/23/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that tip detachment occurred.The patient presented with coronary artery disease and underwent percutaneous coronary intervention.A 185cm pt2 guidewire was selected for use.The wire was attempted to be advanced up the radial artery; however, while advancing the wire, the transition of the floppy tip to the stiffer portion of the wire was fractured.The device was retrieved with a snare.A new pt2 guidewire was then used; however, when the insertion was attempted, the new guidewire tip also fractured.A replacement wire of a different type was used, and the procedure was completed successfully.Following the procedure, while the third pt2 wire was examined in the control room, the tip of the guidewire was fractured.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: unit returned in a generic plastic bag, overall visual revision did not identify failures or evidence that could be lost due to decontamination process.The device was visually and microscopically examined.The proximal section of the wire was kinked; and the distal tip was bent.The reported guidewire fracture is confirmed as the distal section of the wire was found detached.
 
Event or Problem Description
It was reported that tip detachment occurred.The patient presented with coronary artery disease and underwent percutaneous coronary intervention.A 185cm pt2 guidewire was selected for use.The wire was attempted to be advanced up the radial artery; however, while advancing the wire, the transition of the floppy tip to the stiffer portion of the wire was fractured.The device was retrieved with a snare.A new pt2 guidewire was then used; however, when the insertion was attempted, the new guidewire tip also fractured.A replacement wire of a different type was used, and the procedure was completed successfully.Following the procedure, while the third pt2 wire was examined in the control room, the tip of the guidewire was fractured.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PT
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18711576
Report Number2124215-2024-05488
Device Sequence Number7255420
Product Code DQX
UDI-Device Identifier08714729471219
UDI-Public08714729471219
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K030617
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/02/2025
Device Model Number2777
Device Catalogue Number2777
Device Lot Number0030779421
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/23/2024
Supplement Date Received by Manufacturer03/19/2024
Initial Report FDA Received Date02/15/2024
Supplement Report FDA Received Date03/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexFemale
-
-