| Model Number |
K172 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Failure to Capture (1081)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient with this pacemaker device exhibited noisy signals on the right ventricular (rv) channel.The health care professional (hcp) had called in and was questioning if there was loss of capture (loc) as there were couple of fusion markers.The presenting electrogram (egm) showed noise, however it was not sensed.Technical services (ts) stated that there was t-wave post each of the paced indicating there was depolarization.The threshold values were stable.Impedances were fluctuating however they were within the normal range.The plan was to continue monitoring.This device remains in service.No adverse patient effects were reported.
|
| |
|