| Catalog Number |
305195 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Materials#: 305195 batch#: 2363714 it was reported by the customer that coring is occurring when they are inserting in medication vials.Rcc received a complaint via phone.Pir attached.Product complaint-mds facility: (b)(6) ref: 305196 ref 305196 lot: 2nd 3271842 issue: coring is occurring when they are inserting in medication vials.Doe: unknown pt harm: no sample: representative samples available this case is related to (b)(4) when this gets transitioned into tw please relate the cases.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) : initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.The actual date of event is unknown.The date received by manufacturer was entered into the date of event field.
|
| |
|
Additional Manufacturer Narrative
|
|
Pr (b)(4) follow up for device evaluation; it was reported coring is occurring when inserting into medication vials.To aid in the investigation, one sample in a sealed packaging blister was received for evaluation by our quality team.A visual inspection was performed with a 30x microscope.There was no damaged, defective grind or hooks observed.The bevels and etch were good.A device history record review was completed for provided material number 305195, lot 3271842.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.There were no related quality notifications.All processes and final inspections complied with specification requirements.To date, there have been no other similar events reported for this lot.Based on the investigation and with the returned sample analysis the symptom reported by the customer could not be confirmed and a probable root cause could not be determined.
|
| |
|
Event or Problem Description
|
|
No additional information received materials#: 305195 batch#: 3271842 it was reported by the customer that coring is occurring when they are inserting in medication vials.Verbatim: rcc received a complaint via phone.Pir attached.Productcomplaint-mds; facility: (b)(6) hospital.Address: (b)(6).Dept: in patient pharmacy.Contact: 1.(b)(6) iv team lead 2.(b)(6) pharmacy infusion manager phone: 1.(b)(6).2.(b)(6).Email: 1.(b)(6).2.(b)(6).Ref:1st 305196 lot: 1st 2363714 ref 305195 lot: 2nd 3271842 issue: coring is occurring when they are inserting in medication vials.Doe: unknown.Pt harm: no.Sample: representative samples available.This case is related to (b)(4) when this gets transitioned into tw please relate the cases.
|
| |
|