• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WELCH ALLYN INC BT CSM,NIBP,MASIMO,SURETEMP,PLUGB; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WELCH ALLYN INC BT CSM,NIBP,MASIMO,SURETEMP,PLUGB; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS Back to Search Results
Model Number 73MT-B
Medical Device Problem Code Sparking (2595)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The connex spot monitors are intended to be used by clinicians and medically qualified personnel for monitoring of non-invasive blood pressure, pulse rate, non-invasive functional oxygen saturation of arteriolar hemoglobin (spo2), respiration rate, and body temperature in normal and axillary modes of neonatal, pediatric and adult patients.The most likely locations for patients to be monitored are general medical or surgical floors and general hospital and alternate care environments.This product is available for sale only upon the order of a physician or licensed health care professional.The 7000-ps is the connex spot monitor 35 watt power supply hillrom requested for the faulty device to be returned for investigation into the root cause of the reported malfunction however, the device has not been received back from the customer.Although there was no serious injury as a result of this event, if the reported malfunction of sparking and charring of the power supply were to recur, it could result in a serious injury or death, therefore, hillrom is reporting this malfunction.
 
Event or Problem Description
The customer reported the csm does not power on or charge.It was additionally reported the device was plugged into a surge protector and device was sparking where it was plugged in under the device and smelled of burning.There was no adverse event reported.This incident was captured under hillrom complaint ref (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BT CSM,NIBP,MASIMO,SURETEMP,PLUGB
Common Device Name
MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS
Manufacturer (Section D)
WELCH ALLYN INC
4341 state street
skaneateles falls NY 13153
Manufacturer Contact
keighley crosthwaite
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key18712405
Report Number1316463-2024-00023
Device Sequence Number15480529
Product Code MWI
UDI-Device Identifier00732094210071
UDI-Public732094210071
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number73MT-B
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2024
Initial Report FDA Received Date02/15/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-