| Catalog Number |
MS25520 |
| Medical Device Problem Codes |
Loose or Intermittent Connection (1371); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation has just started; results will be provided in a follow-up report.
|
| |
|
Event or Problem Description
|
|
The customer reported the draeger m540 rebooted.It was originally reported as a disconnect, but once on site, it was actually a m540 reboot and came back on.There was no report of an adverse event or patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that an m540 rebooted and came back on during a surgery.The customer replaced the iacs with a delta, and the draeger field service engineer (fse) pulled the logs.The fse installed a new ethernet cable, power supply and system connect board on the defunct ps2500.No additional information regarding how the malfunction was determined was provided.The logs were analyzed; however, the provided logs do not show enough information for the date of event to provide a root cause.The logs confirmed that there was an unexpected reboot on the date of event but cannot provide any detailed information.As neither specific device malfunction or log findings are available, a specific root cause could not be determined.The p2500 was determined to be the cause of the reboot, but the specific defect could not be determined.After repair, the fse performed a full preventive maintenance on the iacs system and when all tests passed, the fse placed the device back into service.
|
| |
|
Event or Problem Description
|
|
The customer reported the draeger m540 rebooted.It was originally reported as a disconnect, but once on site, it was actually a m540 reboot and came back on.There was no report of an adverse event or patient injury.
|
| |
|