• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1012632-29
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Difficult to Advance (2920); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/22/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.In this case, it is likely that the device interacted with the mildly calcified, heavily tortuous, heavily stenosed lesion and previously implanted stents during advancement, as resistance was noted, causing the reported failure to advance.Interaction with the 6f (french) guide catheter, as strong resistance was noted, likely contributed to the reported difficult to advance/position due to the device not being properly supported or coaxially aligned during advancement.Further interaction with the previously implanted stents during retraction of the device likely contributed to the reported difficult to remove, ultimately causing the reported stent dislodgement.The investigation determined the reported difficulties and foreign body in patient, surgical intervention, in addition to unexpected medical intervention (additional therapy/non-surgical treatment) appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily stenosed lesion in the left common iliac artery with mild calcification and heavy tortuosity.There were two previously implanted stents in the external iliac.The 9x29mm otw omni elite balloon-expandable stent (bes) was advanced and strong resistance was noted with the 6f sheath.The bes and the sheath met strong resistance advancing in the anatomy, and the bes failed to cross to the target lesion.The bes was removed from the sheath to determine the cause of the resistance; however, during removal the bes got stuck with the edge of a previously implanted stent, the omni elite stent separated off the balloon into the left common iliac.Attempts were made for 2 hours to snare the stent, but this was unsuccessful and the patient was sent to surgery.The stent was unable to be removed with surgery; therefore, the patient was hospitalized for observation.The stent was later embedded in the arterial wall with a covered stent.There was no adverse patient sequela.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18713035
Report Number2024168-2024-01878
Device Sequence Number8059476
Product Code NIO
UDI-Device Identifier08717648178542
UDI-Public08717648178542
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Catalogue Number1012632-29
Device Lot Number2042041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/22/2024
Initial Report FDA Received Date02/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6FR SHEATH
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-