| Lot Number |
590996 |
| Medical Device Problem Codes |
Difficult to Open or Close (2921); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that during a patient procedure a valve was "stuck" in the scope and could not be actuated.User facility personnel utilized another scope and the procedure was completed successfully.A procedure delay was reported due to exchange of the scopes.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation of this event is currently in process.A followup report will be submitted when additional information becoavmes ailable.
|
| |
|
Additional Manufacturer Narrative
|
|
The user facility returned four (4) unopened/unused valve kits for evaluation.The unopened returned kits were visually inspected, and no issues were noted.The suction valves were functionally tested on a scope to confirm if the reported "sticking" could be duplicated.All four suction valves passed the functional test and performed as designed; no sticking was observed and the reported event could not be duplicated.The valve kit subject of the reported event was not returned for evaluation.Without the return of the subject device, a root cause cannot be determined.The device history record (dhr) was reviewed, and no abnormalities were noted.A complaint review was conducted and confirmed this to be an isolated event.No additional issues have been reported.
|
| |
|