| Catalog Number |
9-ASD-018 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that a 18mm amplatzer septal occluder was selected for implant on (b)(6)2024 using an 8f amplatzer trevisio intravascular delivery system.During implant the left disc of the device took on a snake shape.Another attempt was made to deploy the device.The disc maintained the snake shape.The device was removed from the patient and replaced with a new 19mm amplatzer septal occluder.There were no interactions with cardiac structures.There were no angulations or kinks noted in the delivery system.The patient remained hemodynamically stable throughout the procedure.There were no clinically significant delays in the procedure.There were no adverse effects to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity was reported.A returned device inspection could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Information from the field indicated that there was no anatomical interference, there was no angulation or kink in the delivery system upon deployment and a 8f size delivery system was used.Based on the information received, the cause of the reported device deformity could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that a 18mm amplatzer septal occluder was selected for implant on (b)(6) 2024 using an 8f amplatzer trevisio intravascular delivery system.During implant the left disc of the device took on a snake shape.Another attempt was made to deploy the device.The disc maintained the snake shape.The device was removed from the patient and replaced with a new 19mm amplatzer septal occluder.There were no interactions with cardiac structures.There were no angulations or kinks noted in the delivery system.The patient remained hemodynamically stable throughout the procedure.There were no clinically significant delays in the procedure.There were no adverse effects to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of difficult to fold, unfold or collapse associated with device deformation did not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Information from the field indicated that there was no anatomical interference, there was no angulation or kink in the delivery system upon deployment and a 8f size delivery system was used.Based on the information received, the cause of the reported device deformity could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|
| |
|