| Model Number |
MAF-GM2 |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned, and the evaluation found no malfunctions aside from the allegation reported in b5.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the airway mobilescope connecting tube had peeling coating.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the peeling coating of the connecting tube was caused by stress of repeated use, external factors, or handling of the device.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|