MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD IAG BC PRO GLOBAL PNK 20GA X 1.0IN; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD IAG BC PRO GLOBAL PNK 20GA X 1.0IN; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381033
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd iag bc pro global pnk 20ga x 1.0in had foreign matter.The following information was provided by the initial reporter, translated from japanese to english: this is a complaint about foreign substance contamination.It was reported that when the customer took off the cap, he found a black garbage-like substance on the needle.
 
Additional Manufacturer Narrative
Investigation results: the complaint of foreign matter was confirmed; however, the source of the material could not be determined due to the sample condition.Two photographs and one 20g insyte autoguard were provided for investigation.The catheter was received in open packaging.A dark colored material was found on the catheter.A material analysis showed that the material most closely resembled a plastic material, but the source and exact composition could not be determined.As the device has been opened, it could not be determined with certainty whether the material was introduced during manufacturing or after opening the unit package.A review of the inspection records and quality/manufacturing controls for the implicated lot indicated no issues with the manufacturing process.Manufacturing controls are in place to mitigate the occurrence of this type of failure.The complaint has been documented and will continue to be monitored as part of ongoing efforts to identify potential manufacturing related issues.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD IAG BC PRO GLOBAL PNK 20GA X 1.0IN
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18717024
Report Number1710034-2024-00108
Device Sequence Number3306565
Product Code FOZ
UDI-Device Identifier00382903810338
UDI-Public(01)00382903810338
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/31/2024
Device Catalogue Number381033
Device Lot Number2028733
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2024
Supplement Date Received by Manufacturer04/10/2024
Initial Report FDA Received Date02/15/2024
Supplement Report FDA Received Date04/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-