| Model Number |
RMU-1000 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported that when the rmu-1000 was turned on, there was an alarm light a warning sound and would not function.They reported that this did not occur during patient use.
|
| |
|
Additional Manufacturer Narrative
|
|
Although requested, the device associated with this complaint has not been returned and the cause of the complaint is not known.Should additional information become available a follow-up mdr shall be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the device logs identified that the unit was manufactured in 2019 and had performed a total of (b)(4) compressions.The logs documented intermittent motor-related fault conditions, including "motor didn't move in time" and "stroke depth not within specifications." functional testing confirmed that the device was capable of performing compressions; however, abnormal binding noise was observed during operation.Internal inspection identified damage to the motor connector on the motor control board which did not affect device operation.The exact root cause of the intermittent motor performance degradation could not be conclusively determined based on the available evidence.
|
| |
|