| Catalog Number |
21-7059-24 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that during use, the device over-infused medication.The patient was sent home with 115ml at 2.5ml/hour and the next day he presented to disconnect the pump with the pump showing 10.1ml remaining but the bag was empty.Per the reporter, there were no adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
Two samples were returned for evaluation.Visual inspection revealed no damage or other defects.Functional testing revealed no discrepancies.The failure mode was not confirmed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.D9: date returned to mfg 3/4/2024.
|
| |
|