(b)(4).Method: the complaint opt318 optiflow junior nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Due to this being reported by an anonymous customer, no further information could be provided.Our investigation is based on the information provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Results: the customer stated that the opt318 optiflow junior nasal cannula had been pierced.There was no patient involvement reported by the customer.Conclusion: without the return of the subject device, we are unable to determine the cause of the reported damage.All optiflow junior cannula are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are taken hourly from each run and pull tested to check glue joint strength at the cannula/tube joint, as well as the swivel grip joint, if there are any failures the entire batch is put aside for further investigation.The subject opt318 optiflow junior nasal cannula would have met the required specifications.The user instructions which accompany the opt318 optiflow junior nasal cannula show in pictorial format the correct placement and fitting of the cannula, and also provide the following warnings: "ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient." "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." - "do not crush or stretch tube.".
|